GE LOGIQe: Step 2 - Image Settings

Setup Imaging Preset Settings

  1. Begin new Exam
    1. Select B mode (B key on extended keyboard)
    2. Be sure Harmonics are off by toggling the H key.  
      1. Harmonics are on if the top right setting displays a "CHI" next to the 'B'
    3. Preset key
    4. Select application 'MSK'
    5. Save 'New'
    6. name it 'MuscleSound' 
    7. Overwrite desired preset
    8. Save
    9. Exit
    10. If 'CHI' is displayed, toggle H key, then overwrite the preset
    11. Utility key
      1. Imaging
        1. B
        2. Preset 'MuscleSound'
          1. B/M Acoustic Output %: 100
          2. No Image Reverse
          3. B Color: Tint Map D
          4. Thermal Index: Tls
          5. ATO Level: Medium
          6. Softener: Low
          7. Focus Number: 2
          8. Focus Number CrossXBeam: 2
          9. Focus Width: Wide
          10. Focus Depth %: 52
          11. Depth: 3.5
          12. Compression: 1
          13. Line Density: 3
          14. Line Density Zoom: 3
          15. Suppression: 1
          16. Frame Average: 2
          17. Frame Average CrossXBeam: 1
          18. CrossXBeam: Yes
          19. CrossXBeam #: Low
          20. Compound Type: Mean
          21. Auto Optimize Mode: ATO
          22. Edge Enhance: 3
          23. Width: 1
          24. Gray Scale Map: BC_GRAYD
          25. Gain: 60
          26. Dynamic Range: 66
          27. Virtual Convex: No
          28. Rejection: 0
          29. Frequency: 12
          30. AutoLine Density: No
          31. SRI HD: 3
          32. Gamma: None
          33. Image Size: Small
          34. Image Style: Top

Still need help? Contact Us Contact Us