GE LOGIQe: Step 1 - Connectivity
Setup Connectivity
- DICOM license OR
- an inexpensive workaround that is a viable solution for many (not all) LOGIQe machines.
Wireless USB adapter (Netgear WG111)
- Utilities > Connectivity
- Device
- Select MyComputer
- set or share the AE Title: <device name from musclesound>
- Please communicate with MuscleSound the AE Title your machine was set to, if you don't want to change it.
- Port Number 104
- Add
- Name: DCM4CHEE
- IP Address: 18.205.49.72
- Service
- Destination Device > DCM4CHEE
- Select Service to Add > Dicom Image Storage
- Name: DCM4CHEE
- AE Title: DCM4CHEE
- Port Number: 80
- Maximum Retries: 2
- Retry Interval: 30
- Timeout: 120
- Allow Multiframe: no
- Allow Raw Data: no
- Compression: None
- Quality %: Lossless
- Max Framerate: Full
- Color Support: Mixed
- Reopen Per Image: No
- Enable Structured Reporting: No
- Dataflow
- Add
- Name: DCM4CHEE
- Direct Store: no
- Hidden: no
- Default Dataflow: DCM4CHEE
- Button
- Select Desired Print Button (Print2)
- Format: Dicom (*.dcm)
- Image Frames: Secondary Capture
- Capture Area: Video Area
- Compression: None
- Single Association: no
- Move DCM4CHEE to the Printflow View
- Select child DCM4CHEE
- Select '>>'
Test Connectivity
- Utilities > Connectivity > Service
- Destination Device 'DCM4CHEE'
- Select 'DCM4CHEE' from list at left
- Verify by selecting the yellow face
- Smile Face indicates connection to MuscleSound server
- Sad Face indicates limited or no connection to MuscleSound server
- Verify settings are correct
- Verify connection to local network
- Verify network requirements are met